Breathing performance of regulators


The breathing performance of regulators is a measure of the ability of a breathing gas regulator to meet the demands placed on it at varying ambient pressures and under varying breathing loads, for the range of breathing gases it may be expected to deliver. Performance is an important factor in design and selection of breathing regulators for any application, but particularly for underwater diving, as the range of ambient operating pressures and variety of breathing gases is broader in this application. A diving regulator is a device that reduces the high pressure in a diving cylinder or surface supply hose to the same pressure as the diver's surroundings. It is desirable that breathing from a regulator requires low effort even when supplying large amounts of breathing gas as this is commonly the limiting factor for underwater exertion, and can be critical during diving emergencies. It is also preferable that the gas is delivered smoothly without any sudden changes in resistance while inhaling or exhaling. Although these factors may be judged subjectively, it is convenient to have a standard by which the many different types and manufactures of regulators may be compared.
Various breathing machines have been developed and used for assessment of breathing apparatus performance. Ansti Test Systems developed a turnkey system that measures the inhalation and exhalation effort in using a regulator, and produces graphs indicating the work of breathing at the set depth pressure and respiratory minute volume for the gas mixture used. Publishing results of the performance of regulators in the ANSTI test machine has resulted in performance improvements.

Applications

Breathing performance of the regulator is relevant in all circumstances where a demand regulator is used to provide breathing gas. In some of these applications, a very basic regulator will perform adequately. In other applications, the performance of the regulator may limit the performance of the user. A high-performance regulator for a given combination of gas mixture and ambient pressure will provide a low work of breathing at high RMV.
A healthy person at rest at surface atmospheric pressure expends only a small amount of available effort on breathing. This can change considerably as the density of breathing gas increases at higher ambient pressure. When the energy expended to remove carbon dioxide produces more carbon dioxide than it removes the person will suffer from hypercapnia in a positive feedback cycle ending in unconsciousness and eventually death. Work of breathing is affected by breathing rate, breathing pattern, gas density, physiological factors, and the fluid dynamic details of the breathing apparatus, the namely frictional resistance to flow, and pressure differences required to open valves and hold them open to flow.
Breathing gas density can be reduced by using helium as the basic component, with sufficient oxygen added to suit the circumstances and retain a partial pressure sufficient to sustain consciousness but not so much as to cause oxygen toxicity problems. Frictional resistance to flow is influenced by the shape and size of the gas passages, and the pressure, density, viscosity, and velocity of the gas. Valve cracking pressure is a factor of design and settings of the valve mechanisms. The breathing performance of regulators assumes gas density is specified and measures the resistance to flow during the full breathing cycle with a given volumetric flow rate as a pressure drop between the mouthpiece and the exterior environment.

Measurement

Work of breathing

Work of breathing is the energy expended to inhale and exhale a breathing gas. It is usually expressed as work per unit volume, for example, joules/litre, or as a work rate, such as joules/min or equivalent units, as it is not particularly useful without a reference to volume or time. It can be calculated in terms of the pulmonary pressure multiplied by the change in pulmonary volume, or in terms of the oxygen consumption attributable to breathing.
The total work of breathing when using a breathing apparatus is the sum of the physiological work of breathing and the mechanical work of breathing of the apparatus.
In a normal resting state the physiological work of breathing constitutes about 5% of the total body oxygen consumption. It can increase considerably due to illness or constraints on gas flow imposed by breathing apparatus, ambient pressure, or breathing gas composition.

ANSTI machine

The ANSTI Breathing Simulator is rated to a maximum working pressure of 100 msw. It uses a piston mechanism to provide an accurate and repeatable volume displacement with a sine wave drive mechanism. It has adjustable tidal volume and breathing rate settings which can providing ventilation rates from 10 to 180 litres per minute.

EU Standards

In the European Union the standard EN250:2000 Respiratory equipment. Open-circuit self-contained compressed air diving apparatus. Requirements, testing, marking defines minimum performance standards for "Open-circuit self-contained compressed air diving apparatus", and BS 8547:2016 defines requirements for demand regulators to be used at depths exceeding 50 m. EN 13949: 2003 – Respiratory Equipment – Open Circuit Self-Contained Diving Apparatus for use with Compressed Nitrox and Oxygen – Requirements, Testing, Marking defines requirements for regulators to be used with raised levels of oxygen.
The standard contains limits on inhalation and exhalation pressures and overall work of breathing. It specifies the following, under test conditions of a breathing rate of per minute and an ambient pressure of :
Although a regulator meeting the above limits will supply sufficient air where the first stage feeds a single second stage, it is not necessarily capable of supplying sufficient air in all circumstances when a single first stage feeds two second stages simultaneously.

Related standards

In Europe, EN 250: 2014 – Respiratory Equipment – Open Circuit Self - Contained Compressed
Air Diving Apparatus – Requirements, Testing and Marking defines the minimum requirements for breathing performance of regulators, and BS 8547:2016 defines requirements for demand regulators to be used at depths exceeding 50 m. EN 13949: 2003 – Respiratory Equipment – Open Circuit Self-Contained Diving Apparatus for use with Compressed Nitrox and Oxygen – Requirements, Testing, Marking. defines requirements for regulators to be used with raised levels of oxygen.
EN 15333 – 1: 2008 COR 2009 – Respiratory Equipment – Open-Circuit Umbilical Supplied Compressed Gas Diving Apparatus – Part 1: Demand Apparatus. and EN 15333 – 2: 2009 – Respiratory Equipment – Open-Circuit Umbilical Supplied Compressed Gas Diving Apparatus – Part 2: Free Flow Apparatus.
I.S. EN 14143: 2013 – Respiratory Equipment – Self-Contained Re-Breathing Diving Apparatus defines minimum requirements for rebreathers.

US Military

In the United States Military, the standard for single-hose scuba regulators was MIl-R-24169B, now withdrawn.