Charged aerosol detector
The Charged Aerosol Detector is a detector used in conjunction with high-performance liquid chromatography and ultra high-performance liquid chromatography to measure the amount of chemicals in a sample by creating charged aerosol particles which are detected using an electrometer. It is commonly used for the analysis of compounds that cannot be detected using traditional UV/Vis approaches due to their lack of a chromophore. The CAD can measure all non-volatile and many semi-volatile analytes including, but not limited to, antibiotics, excipients, ions, lipids, natural products, biofuels, sugars and surfactants. The CAD, like other aerosol detectors and condensation nucleation light scattering detectors ), falls under the category of destructive general-purpose detectors.
History
The predecessor to the CAD, termed an evaporative electrical detector, was first described by Kaufman at in and was based on the coupling of liquid chromatographic approaches to TSI’s electrical aerosol measurement technology. At around the same time Dixon and Peterson at California State University were investigating the coupling of liquid chromatography to an earlier version of TSI’s EAM technology, which they called an aerosol charge detector. Subsequent collaboration between TSI and ESA Biosciences Inc., led to the first commercial instrument, the Corona CAD, which received both the Pittsburgh Conference Silver Pittcon Editor’s Award and . Continued research and engineering improvements in product design resulted in . The newest iterations of the CAD are the and and2005 | 2006 | 2009 | 2011 | 2013 | 2015 |
ESA Biosciences, Inc. Corona CAD | ESA Biosciences, Inc. Corona PLUS | ESA Biosciences, Inc. Corona ultra | Dionex Corona ultra RS | Dionex Corona Veo | Vanquish Charged Aerosol Detector |
•First commercial CAD •Designed for near-universal detection on any HPLC •Isocratic or gradient separations | •Expanded solvent compatibility •Heated nebulization •External gas conditioning module for improved precision | •UHPLC compatible •Stackable design •Enhanced sensitivity •Incorporated precision internal gas regulation system | •Unified with Dionex UltiMate 3000 UHPLC+ system •Added on-board diagnostics/monitoring •Automated flow diversion capability •Selection of linearization parameters | •Extended micro flow rate range •Total redesign with concentric nebulization and optimized spray chamber •Heated evaporation and electronic gas regulation | •Full integration with Thermo Scientific Vanquish UHPLC platform •Slide-in module design •Reduced flow path for optimum operation |
Principles of Operation
The involves:- Pneumatic nebulization of mobile phase from the analytical column forming an aerosol.
- Aerosol conditioning to remove large droplets.
- Evaporation of solvent from the droplets to form dried particles.
- Particle charging using an ion jet formed via corona discharge.
- Particle selection - an ion trap is used to excess ions and high mobility charged particles.
- Measurement of the aggregate charge of aerosol particles using a filter/electrometer.
More detailed information on how CAD works can be found on the .
CAD Performance and Comparison to Other Aerosol Detectors
- The CAD and evaporative light scattering detector are mass-flow sensitive detectors as opposed to concentration sensitive detectors such as UV detectors.
- Both the CAD and ELSD exhibit non-linear responses for most sample types, however over small ranges CAD response is reasonably linear. The shape of the response curves are different between the two detectors.
- Both detectors require the use of fully volatile mobile phases and non-volatile samples. CAD response is dependent on organic content of the mobile phase, response is higher with organic-rich mobile phases than aqueous ones.
- CAD response is quite uniform for non-volatile analytes with detection limit 1 - 3ng, however response for ionised basic analytes can be larger than for neutral analytes.
- When properly individually optimized, both the CAD and ELSD show similar responses.
- Failure to regularly clean and optimize the mobile phase and gas flow properties of the detector leads to intra- and inter-day precision/ reproducibility errors.