Charged aerosol detector


The Charged Aerosol Detector is a detector used in conjunction with high-performance liquid chromatography and ultra high-performance liquid chromatography to measure the amount of chemicals in a sample by creating charged aerosol particles which are detected using an electrometer. It is commonly used for the analysis of compounds that cannot be detected using traditional UV/Vis approaches due to their lack of a chromophore. The CAD can measure all non-volatile and many semi-volatile analytes including, but not limited to, antibiotics, excipients, ions, lipids, natural products, biofuels, sugars and surfactants. The CAD, like other aerosol detectors and condensation nucleation light scattering detectors ), falls under the category of destructive general-purpose detectors.

History

The predecessor to the CAD, termed an evaporative electrical detector, was first described by Kaufman at in and was based on the coupling of liquid chromatographic approaches to TSI’s electrical aerosol measurement technology. At around the same time Dixon and Peterson at California State University were investigating the coupling of liquid chromatography to an earlier version of TSI’s EAM technology, which they called an aerosol charge detector. Subsequent collaboration between TSI and ESA Biosciences Inc., led to the first commercial instrument, the Corona CAD, which received both the Pittsburgh Conference Silver Pittcon Editor’s Award and . Continued research and engineering improvements in product design resulted in . The newest iterations of the CAD are the and and
200520062009201120132015
ESA Biosciences, Inc.
Corona
CAD
ESA Biosciences, Inc.
Corona
PLUS
ESA Biosciences, Inc.
Corona
ultra
Dionex
Corona
ultra RS

Dionex
Corona
Veo

Vanquish
Charged Aerosol Detector
•First commercial CAD
•Designed for near-universal
detection on any HPLC
•Isocratic or gradient
separations
•Expanded solvent compatibility
•Heated nebulization
•External gas conditioning
module for improved precision
•UHPLC compatible
•Stackable design
•Enhanced sensitivity
•Incorporated precision
internal gas regulation
system
•Unified with Dionex
UltiMate 3000 UHPLC+
system
•Added on-board
diagnostics/monitoring
•Automated flow
diversion capability
•Selection of linearization
parameters
•Extended micro flow
rate range
•Total redesign with
concentric nebulization
and optimized spray
chamber
•Heated evaporation
and electronic gas
regulation
•Full integration with Thermo
Scientific Vanquish
UHPLC platform
•Slide-in module design
•Reduced flow path for
optimum operation

Principles of Operation

The involves:
The CAD like other aerosol detectors, can only be used with volatile mobile phases. For an analyte to be detected it must be less volatile than the mobile phase.
More detailed information on how CAD works can be found on the .

CAD Performance and Comparison to Other Aerosol Detectors