ELISA
The enzyme-linked immunosorbent assay is a commonly used analytical biochemistry assay, first described by Engvall and Perlmann in 1971. The assay uses a solid-phase type of enzyme immunoassay to detect the presence of a ligand in a liquid sample using antibodies directed against the protein to be measured. ELISA has been used as a diagnostic tool in medicine, plant pathology, and biotechnology, as well as a quality control check in various industries.
In the most simple form of an ELISA, antigens from the sample to be tested are attached to a surface. Then, a matching antibody is applied over the surface so it can bind the antigen. This antibody is linked to an enzyme and then any unbound antibodies are removed. In the final step, a substance containing the enzyme's substrate is added. If there was binding the subsequent reaction produces a detectable signal, most commonly a color change.
Performing an ELISA involves at least one antibody with specificity for a particular antigen. The sample with an unknown amount of antigen is immobilized on a solid support either non-specifically or specifically. After the antigen is immobilized, the detection antibody is added, forming a complex with the antigen. The detection antibody can be covalently linked to an enzyme or can itself be detected by a secondary antibody that is linked to an enzyme through bioconjugation. Between each step, the plate is typically washed with a mild detergent solution to remove any proteins or antibodies that are non-specifically bound. After the final wash step, the plate is developed by adding an enzymatic substrate to produce a visible signal, which indicates the quantity of antigen in the sample.
Of note, ELISA can perform other forms of ligand binding assays instead of strictly "immuno" assays, though the name carried the original "immuno" because of the common use and history of development of this method. The technique essentially requires any ligating reagent that can be immobilized on the solid phase along with a detection reagent that will bind specifically and use an enzyme to generate a signal that can be properly quantified. In between the washes, only the ligand and its specific binding counterparts remain specifically bound or "immunosorbed" by antigen-antibody interactions to the solid phase, while the nonspecific or unbound components are washed away. Unlike other spectrophotometric wet lab assay formats where the same reaction well can be reused after washing, the ELISA plates have the reaction products immunosorbed on the solid phase, which is part of the plate, and so are not easily reusable.
Principle
As an analytical biochemistry assay and a "wet lab" technique, ELISA involves detection of an analyte in a liquid sample by a method that continues to use liquid reagents during the analysis that stays liquid and remains inside a reaction chamber or well needed to keep the reactants contained. This is in contrast to "dry lab" techniques that use dry strips. Even if the sample is liquid, the final detection step in "dry" analysis involves reading of a dried strip by methods such as reflectometry and does not need a reaction containment chamber to prevent spillover or mixing between samples.As a heterogenous assay, ELISA separates some component of the analytical reaction mixture by adsorbing certain components onto a solid phase which is physically immobilized. In ELISA, a liquid sample is added onto a stationary solid phase with special binding properties and is followed by multiple liquid reagents that are sequentially added, incubated, and washed, followed by some optical change in the final liquid in the well from which the quantity of the analyte is measured. The quantitative "reading" is usually based on detection of intensity of transmitted light by spectrophotometry, which involves quantitation of transmission of some specific wavelength of light through the liquid. The sensitivity of detection depends on amplification of the signal during the analytic reactions. Since enzyme reactions are very well known amplification processes, the signal is generated by enzymes which are linked to the detection reagents in fixed proportions to allow accurate quantification, and thus the name "enzyme-linked."
The analyte is also called the ligand because it will specifically bind or ligate to a detection reagent, thus ELISA falls under the bigger category of ligand binding assays. The ligand-specific binding reagent is "immobilized," i.e., usually coated and dried onto the transparent bottom and sometimes also side wall of a well, which is usually constructed as a multiple-well plate known as the "ELISA plate." Conventionally, like other forms of immunoassays, the specificity of antigen-antibody type reaction is used because it is easy to raise an antibody specifically against an antigen in bulk as a reagent. Alternatively, if the analyte itself is an antibody, its target antigen can be used as the binding reagent.
History
Before the development of the ELISA, the only option for conducting an immunoassay was radioimmunoassay, a technique using radioactively labeled antigens or antibodies. In radioimmunoassay, the radioactivity provides the signal, which indicates whether a specific antigen or antibody is present in the sample. Radioimmunoassay was first described in a scientific paper by Rosalyn Sussman Yalow and Solomon Berson published in 1960.As radioactivity poses a potential health threat, a safer alternative was sought. A suitable alternative to radioimmunoassay would substitute a nonradioactive signal in place of the radioactive signal. When enzymes react with appropriate substrates, a change in color occurs, which is used as a signal. However, the signal has to be associated with the presence of antibody or antigen, which is why the enzyme has to be linked to an appropriate antibody. This linking process was independently developed by Stratis Avrameas and G. B. Pierce. Since it is necessary to remove any unbound antibody or antigen by washing, the antibody or antigen has to be fixed to the surface of the container; i.e., the immunosorbent must be prepared. A technique to accomplish this was published by Wide and Jerker Porath in 1966.
In 1971, Peter Perlmann and Eva Engvall at Stockholm University in Sweden, and Anton Schuurs and Bauke van Weemen in the Netherlands independently published papers that synthesized this knowledge into methods to perform EIA/ELISA.
Traditional ELISA typically involves chromogenic reporters and substrates that produce some kind of observable color change to indicate the presence of antigen or analyte. Newer ELISA-like techniques use fluorogenic, electrochemiluminescent, and quantitaoppositiontive PCR reporters to create quantifiable signals. These new reporters can have various advantages, including higher sensitivities and multiplexing. In technical terms, newer assays of this type are not strictly ELISAs, as they are not "enzyme-linked", but are instead linked to some nonenzymatic reporter. However, given that the general principles in these assays are largely similar, they are often grouped in the same category as ELISAs.
In 2012, an ultrasensitive, enzyme-based ELISA test using nanoparticles as a chromogenic reporter was able to give a naked-eye colour signal, from the detection of mere attograms of analyte. A blue color appears for positive results and red color for negative. Note that this detection only can confirm the presence or the absence of analyte, not the actual concentration.
Types
There are many ELISA tests for particular molecules that use the matching antibodies. ELISA tests are broken into several types of tests based on how the analytes and antibodies are bonded and used. The major types are described here.Direct ELISA
The steps of direct ELISA follows the mechanism below:- A buffered solution of the antigen to be tested for is added to each well of a microtiter plate, where it is given time to adhere to the plastic through charge interactions.
- A solution of nonreacting protein, such as bovine serum albumin or casein, is added to each well in order to cover any plastic surface in the well which remains uncoated by the antigen.
- The primary antibody with an attached enzyme is added, which binds specifically to the test antigen coating the well.
- A substrate for this enzyme is then added. Often, this substrate changes color upon reaction with the enzyme.
- The higher the concentration of the primary antibody present in the serum, the stronger the color change. Often, a spectrometer is used to give quantitative values for color strength.
ELISA may be run in a qualitative or quantitative format. Qualitative results provide a simple positive or negative result for a sample. The cutoff between positive and negative is determined by the analyst and may be statistical. Two or three times the standard deviation is often used to distinguish positive from negative samples. In quantitative ELISA, the optical density of the sample is compared to a standard curve, which is typically a serial dilution of a known-concentration solution of the target molecule. For example, if a test sample returns an OD of 1.0, the point on the standard curve that gave OD = 1.0 must be of the same analyte concentration as the sample.
The use and meaning of the names "indirect ELISA" and "direct ELISA" differs in the literature and on web sites depending on the context of the experiment. When the presence of an antigen is analyzed, the name "direct ELISA" refers to an ELISA in which only a labelled primary antibody is used, and the term "indirect ELISA" refers to an ELISA in which the antigen is bound by the primary antibody which then is detected by a labeled secondary antibody. In the latter case a sandwich ELISA is clearly distinct from an indirect ELISA. When the "primary" antibody is of interest, e.g. in the case of immunization analyses, this antibody is directly detected by the secondary antibody and the term "indirect ELISA" applies to a setting with two antibodies.
Sandwich ELISA
A "sandwich" ELISA is used to detect sample antigen. The steps are:- A surface is prepared to which a known quantity of capture antibody is bound.
- Any nonspecific binding sites on the surface are blocked.
- The antigen-containing sample is applied to the plate, and captured by antibody.
- The plate is washed to remove unbound antigen.
- A specific antibody is added, and binds to antigen. This primary antibody could also be in the serum of a donor to be tested for reactivity towards the antigen.
- Enzyme-linked secondary antibodies are applied as detection antibodies that also bind specifically to the antibody's Fc region.
- The plate is washed to remove the unbound antibody-enzyme conjugates.
- A chemical is added to be converted by the enzyme into a color or fluorescent or electrochemical signal.
- The absorbance or fluorescence or electrochemical signal of the plate wells is measured to determine the presence and quantity of antigen.
Competitive ELISA
A third use of ELISA is through competitive binding. The steps for this ELISA are somewhat different from the first two examples:Unlabeled antibody is incubated in the presence of its antigen.
- These bound antibody/antigen complexes are then added to an antigen-coated well.
- The plate is washed, so unbound antibodies are removed.
- The secondary antibody, specific to the primary antibody, is added. This second antibody is coupled to the enzyme.
- A substrate is added, and remaining enzymes elicit a chromogenic or fluorescent signal.
- The reaction is stopped to prevent eventual saturation of the signal.
Commonly, the antigen is not first positioned in the well.
For the detection of HIV antibodies, the wells of microtiter plate are coated with the HIV antigen. Two specific antibodies are used, one conjugated with enzyme and the other present in serum. Cumulative competition occurs between the two antibodies for the same antigen, causing a stronger signal to be seen. Sera to be tested are added to these wells and incubated at 37 °C, and then washed. If antibodies are present, the antigen-antibody reaction occurs. No antigen is left for the enzyme-labelled specific HIV antibodies. These antibodies remain free upon addition and are washed off during washing. Substrate is added, but there is no enzyme to act on it, so a positive result shows no color change.
Reverse ELISA
A fourth ELISA test does not use the traditional wells. This test leaves the antigens suspended in the test fluid.- Unlabeled antibody is incubated in the presence of its antigen
- A sufficient incubation period is provided to allow the antibodies to bind to the antigens.
- The sample is then passed through the Scavenger container. This can be a test tube or a specifically designed flow through channel. The surface of the Scavenger container or channel has “Scavenger Antigens” bound to it. These can be identical or sufficiently similar to the primary antigens that the free antibodies will bind.
- The Scavenger container must have sufficient surface area and sufficient time to allow the Scavenger Antigens to bind to all the excess Antibodies introduced into the sample.
- The sample, that now contains the tagged and bound antibodies, is passed through a detector. This device can be a flow cytometer or other device that illuminates the tags and registers the response.
This test is done, generally, one test at a time and cannot be done with the microtiter plate. The equipment needed is usually less complicated and can be used in the field.
Commonly used enzymatic markers
The following table lists the enzymatic markers commonly used in ELISA assays, which allow the results of the assay to be measured upon completion.- OPD turns amber to detect HRP, which is often used to as a conjugated protein.
- TMB turns blue when detecting HRP and turns yellow after the addition of sulfuric or phosphoric acid.
- ABTS turns green when detecting HRP.
- PNPP turns yellow when detecting alkaline phosphatase.
Applications
The ELISA was the first screening test widely used for HIV because of its high sensitivity. In an ELISA, a person's serum is diluted 400 times and applied to a plate to which HIV antigens are attached. If antibodies to HIV are present in the serum, they may bind to these HIV antigens. The plate is then washed to remove all other components of the serum. A specially prepared "secondary antibody"—an antibody that binds to other antibodies—is then applied to the plate, followed by another wash. This secondary antibody is chemically linked in advance to an enzyme.
Thus, the plate will contain enzyme in proportion to the amount of secondary antibody bound to the plate. A substrate for the enzyme is applied, and catalysis by the enzyme leads to a change in color or fluorescence. ELISA results are reported as a number; the most controversial aspect of this test is determining the "cut-off" point between a positive and a negative result.
A cut-off point may be determined by comparing it with a known standard. If an ELISA test is used for drug screening at workplace, a cut-off concentration, 50 ng/ml, for example, is established, and a sample containing the standard concentration of analyte will be prepared. Unknowns that generate a stronger signal than the known sample are "positive." Those that generate weaker signal are "negative".
Dr Dennis E Bidwell and Alister Voller created the ELISA test to detect various kind of diseases, such as dengue, malaria, Chagas disease, Johne's disease, and others. ELISA tests also are used as in in vitro diagnostics in medical laboratories. The other uses of ELISA include:
- detection of Mycobacterium antibodies in tuberculosis
- detection of rotavirus in feces
- detection of hepatitis B markers in serum
- detection of hepatitis C markers in serum
- detection of enterotoxin of E. coli in feces
- detection of HIV antibodies in blood samples