Laser-induced breakdown spectroscopy


Laser-induced breakdown spectroscopy is a type of atomic emission spectroscopy which uses a highly energetic laser pulse as the excitation source. The laser is focused to form a plasma, which atomizes and excites samples. The formation of the plasma only begins when the focused laser achieves a certain threshold for optical breakdown, which generally depends on the environment and the target material. In principle, LIBS can analyze any matter regardless of its physical state, be it solid, liquid or gas. Because all elements emit light of characteristic frequencies when excited to sufficiently high temperatures, LIBS can the plasma excitation temperature, b) the light collection window, and c) the line strength of the viewed transition. LIBS makes use of optical emission spectrometry and is to this extent very similar to arc/spark emission spectroscopy.
LIBS operates by focusing the laser onto a small area at the surface of the specimen; when the laser is discharged it ablates a very small amount of material, in the range of nanograms to picograms, which generates a plasma plume with temperatures in excess of 100,000 K. During data collection, typically after local thermodynamic equilibrium is established, plasma temperatures range from 5,000–20,000 K. At the high temperatures during the early plasma, the ablated material dissociates into excited ionic and atomic species. During this time, the plasma emits a continuum of radiation which does not contain any useful information about the species present, but within a very small timeframe the plasma expands at supersonic velocities and cools. At this point the characteristic atomic emission lines of the elements can be observed. The delay between the emission of continuum radiation and characteristic radiation is in the order of 10 μs, which is why it is necessary to temporally gate the detector.
LIBS is sometimes referred to as laser-induced plasma spectroscopy ; however that acronym also has alternative meanings that are outside the field of analytical spectroscopy.
LIBS is technically very similar to a number of other laser-based analytical techniques, sharing much of the same hardware. These techniques are the vibrational spectroscopic technique of Raman spectroscopy, and the fluorescence spectroscopic technique of laser-induced fluorescence. In fact devices are now being manufactured which combine these techniques in a single instrument, allowing the atomic, molecular and structural characterisation of a specimen as well as giving a deeper insight into physical properties.

Design

A typical LIBS system consists of a solid-state laser and a spectrometer with a wide spectral range and a high sensitivity, fast response rate, time gated detector. This is coupled to a computer which can rapidly process and interpret the acquired data. As such LIBS is one of the most experimentally simple spectroscopic analytical techniques, making it one of the cheapest to purchase and to operate.
The Nd:YAG laser generates energy in the near infrared region of the electromagnetic spectrum, with a wavelength of 1064 nm. The pulse duration is in the region of 10 ns generating a power density which can exceed 1 GW·cm−2 at the focal point. Other lasers have been used for LIBS, mainly the Excimer type that generates energy in the visible and ultraviolet regions.
The spectrometer consists of either a monochromator or a polychromator and a photomultiplier or CCD detector respectively. The most common monochromator is the Czerny–Turner type whilst the most common polychromator is the Echelle type. However, even the Czerny-Turner type can be used to disperse the radiation onto a CCD effectively making it a polychromator. The polychromator spectrometer is the type most commonly used in LIBS as it allows simultaneous acquisition of the entire wavelength range of interest.
The spectrometer collects electromagnetic radiation over the widest wavelength range possible, maximising the number of emission lines detected for each particular element. Spectrometer response is typically from 1100 nm to 170 nm, the approximate response range of a CCD detector. All elements have emission lines within this wavelength range. The energy resolution of the spectrometer can also affect the quality of the LIBS measurement, since high resolution systems can separate spectral emission lines in close juxtaposition, reducing interference and increasing selectivity. This feature is particularly important in specimens which have a complex matrix, containing a large number of different elements. Accompanying the spectrometer and detector is a delay generator which accurately gates the detector's response time, allowing temporal resolution of the spectrum.

Advantages

Because such a small amount of material is consumed during the LIBS process the technique is considered essentially non-destructive or minimally-destructive, and with an average power density of less than one watt radiated onto the specimen there is almost no specimen heating surrounding the ablation site.
Due to the nature of this technique sample preparation is typically minimised to homogenisation or is often unnecessary where heterogeneity is to be investigated or where a specimen is known to be sufficiently, this reduces the possibility of contamination during chemical preparation steps. One of the major advantages of the LIBS technique is its ability to depth profile a specimen by repeatedly discharging the laser in the same position, effectively going deeper into the specimen with each shot. This can also be applied to the removal of surface contamination, where the laser is discharged a number of times prior to the analysing shot. LIBS is also a very rapid technique giving results within seconds, making it particularly useful for high volume analyses or on-line industrial monitoring.
LIBS is an entirely optical technique, therefore it requires only optical access to the specimen. This is of major significance as fiber optics can be employed for remote analyses. And being an optical technique it is non-invasive, non-contact and can even be used as a stand-off analytical technique when coupled to appropriate telescopic apparatus. These attributes have significance for use in areas from hazardous environments to space exploration. Additionally LIBS systems can easily be coupled to an optical microscope for micro-sampling adding a new dimension of analytical flexibility.
With specialised optics or a mechanically positioned specimen stage the laser can be scanned over the surface of the specimen allowing spatially resolved chemical analysis and the creation of 'elemental maps'. This is very significant as chemical imaging is becoming more important in all branches of science and technology.
Portable LIBS systems are more sensitive, faster and can detect a wider range of elements than competing techniques such as portable x-ray fluorescence. And LIBS does not use ionizing radiation to excite the sample, which is both penetrating and potentially carcinogenic.

Disadvantages

LIBS, like all other analytical techniques is not without limitations. It is subject to variation in the laser spark and resultant plasma which often limits reproducibility. The accuracy of LIBS measurements is typically better than 10% and precision is often better than 5%. The detection limits for LIBS vary from one element to the next depending on the specimen type and the experimental apparatus used. Even so, detection limits of 1 to 30 ppm by mass are not uncommon, but can range from to

2000s developments

From 2000–2010, the U.S. Army Research Laboratory researched potential extensions to LIBS technology, which focused on hazardous material detection. Applications investigated at ARL included the standoff detection of explosive residues and other hazardous materials, plastic landmine discrimination, and material characterization of various metal alloys and polymers. Results presented by ARL suggest that LIBS may be able to discriminate between energetic and non-energetic materials.

Research

In 2000, ARL and Ocean Optics Inc. developed a broadband high-resolution spectrometer which was commercialized in 2003. Designed for material analysis, the spectrometer allowed the LIBS system to be sensitive to chemical elements in low concentration.
ARL LIBS applications studied from 2000 to 2010 included:  
ARL LIBS prototypes studied during this period included:
In the 2010s, interest developed in LIBS that focused on the miniaturization of the components and the development of compact, low-power, portable systems. Interest from groups such as NASA and ESA - as well as the military - has furthered these developments. The Mars Science Laboratory mission brought ChemCam, a LIBS instrument, to the surface of Mars in 2012.
Recent developments in LIBS have seen the introduction of double-pulsed laser systems. For double-pulse LIBS one distinguishes between orthogonal and perpendicular configuration. In perpendicular configuration the laser fires twice on the same spot on the specimen with a pulse separation in the order of one to a couple of tens of microseconds. Depending on pulse separation, the second pulse is more or less absorbed by the plasma plume caused by the previous pulse, resulting in a reheating of the laser plasma leading to signal enhancement.
In orthogonal configuration a laser pulse is fired parallel to the sample surface either before or after the perpendicular pulse hits the specimen. The laser plasma ignited in the surrounding medium above the surface by a first pulse causes an area of reduced pressure above the specimen into which the actual plasma from the sample can expand. This has similar positive effects on sensitivity like LIBS performed at reduced pressures.
If the orthogonal laser pulse is delayed with respect to the perpendicular one, the effects are similar as in the perpendicular configuration. Timing electronics such as digital delay generators can precisely control the timing of both pulses.
Both double-pulse LIBS as well as LIBS at reduced pressures aim at increasing the sensitivity of LIBS and the reduction of errors caused by the differential volatility of elements. It also significantly reduces the matrix effects. Double-pulsed systems have proven useful in conducting analysis in liquids, as the initial laser pulse forms a cavity bubble in which the second pulse acts on the evaporated material.
LIBS is one of several analytical techniques that can be deployed in the field as opposed to pure laboratory techniques e.g. spark OES., recent research on LIBS focuses on compact and portable systems. Some industrial applications of LIBS include the detection of material mix-ups, analysis of inclusions in steel, analysis of slags in secondary metallurgy, analysis of combustion processes, and high-speed identification of scrap pieces for material-specific recycling tasks. Armed with data analysis techniques, this technique is being extended to pharmaceutical samples.

LIBS using short laser pulses

Following multiphoton or tunnel ionization the electron is being accelerated by inverse Bremsstrahlung and can collide with the nearby molecules and generate new electrons through collisions. If the pulse duration is long, the newly ionized electrons can be accelerated and eventually avalanche or cascade ionization follows. Once the density of the electrons reaches a critical value, breakdown occurs and high density plasma is created which has no memory of the laser pulse. So, the criterion for the shortness of a pulse in dense media is as follows: A pulse interacting with a dense matter is considered to be short if during the interaction the threshold for the avalanche ionization is not reached. At the first glance this definition may appear to be too limiting. Fortunately, due to the delicately balanced behavior of the pulses in dense media, the threshold cannot be reached easily. The phenomenon responsible for the balance is the intensity clamping through the onset of filamentation process during the propagation of strong laser pulses in dense media.
A potentially important development to LIBS involves the use of a short laser pulse as a spectroscopic source. In this method, a plasma column is created as a result of focusing ultrafast laser pulses in a gas. The self-luminous plasma is far superior in terms of low level of continuum and also smaller line broadening. This is attributed to the lower density of the plasma in the case of short laser pulses due to the defocusing effects which limits the intensity of the pulse in the interaction region and thus prevents further multiphoton/tunnel ionization of the gas.

Line intensity

For an optically thin plasma composed of a single, neutral atomic species in local thermal equilibrium, the density of photons emitted by a transition from level i to level j is
where :
The partition function is the statistical occupation fraction of every level of the atomic species :

LIBS for food analysis

Recently, LIBS has been investigated as a fast, micro-destructive food analysis tool. It is considered a potential analytical tool for qualitative and quantitative chemical analysis, making it suitable as a PAT or portable tool. Milk, bakery products, tea, vegetable oils, water, cereals, flour, potatoes, palm date and different types of meat have been analyzed using LIBS. Few studies have shown its potential as an adulteration detection tool for certain foods. LIBS has also been evaluated as a promising elemental imaging technique in meat.
In 2019, researchers of the University of York and of the Liverpool John Moores University employed LIBS for studying 12 European oysters from the Late Mesolithic shell midden at Conors Island. The results highlighted the applicability of LIBS to determine prehistoric seasonality practices as well as biological age and growth at an improved rate and reduced cost than was previously achievable.