Schlieren imaging


Schlieren imaging is a method to visualize density variations in transparent media.
The term "schlieren imaging" is commonly used as a synonym for schlieren photography, though this article particularly treats visualization of the pressure field produced by ultrasonic transducers, generally in water or tissue-mimicking media. The method provides a two-dimensional projection image of the acoustic beam in real-time.
The unique properties of the method enable the investigation of specific features of the acoustic field, detection of acoustic beam-profile irregularities and on-line identification of time-dependent phenomena . Some researchers say that schlieren imaging is equivalent to an X-ray radiograph of the acoustic field.

Setup

The optical setup of a schlieren imaging system may comprise the following main sections:
Parallel beam, focusing element, stop and a camera.
The parallel beam may be achieved by a point-like light source placed in the focal point of a collimating optical element.
The focusing element may be a lens or a mirror.
The optical stop may be realized by a razor placed horizontally or vertically in the focal point of the focusing element, carefully positioned to block the light spot image on its edge.
The camera is positioned behind the stop and may be equipped with a suitable lens.

Physics

Ray optics description

A parallel beam is described as a group of straight and parallel 'rays'.
The rays cross through the transparent medium while potentially interacting with the contained acoustic field, and finally reach the focusing element.
Note that the principle of a focusing element is directing rays that are parallel - into a single point on the focal plane of the element.
Thus, the population of rays crossing the focal plane of the focusing element can be divided into two groups: those that interacted with the acoustic field and those that didn't. The latter group is undisturbed by the acoustic field, so it remains parallel and forms a point in a well-defined position in the focal plane. The optical stop is positioned exactly at that point, so as to prevent all corresponding rays from further propagating through the system and to the camera.
Thus we get rid of the portion of light that crossed the acoustic field without interaction.
However, there are also rays that did interact with the acoustic field in the following manner:
If a ray travels through a region of nonuniform density whose spatial gradient has a component orthogonal to the ray, that ray is deflected from its original orientation, as if it were passing through a prism. This ray is no longer parallel, so it doesn't intersect the focal point of the focusing element and is not blocked by the knife. In some circumstances the deflected ray escapes the knife-blade and reaches the camera to create a point-like image on the camera-sensor, with a position and intensity related to the inhomogeneity experienced by the ray. An image is formed in this way, exclusively by rays that interacted with the acoustic field, providing a mapping of the acoustic field.

Physical optics description

The acousto-optic effect couples the optical refractive index of the medium with its density and pressure. Thus, spatial and temporal variations in pressure induces corresponding variations in refractive index. Optical wavelength and wavenumber in medium depend on refractive index. The phase acquired by electromagnetic wave traveling through the medium is related to the line-integral of the wavenumber along the propagation line.
For a plane-wave electromagnetic radiation traveling parallel to the Z-axis, the XY planes are iso-phase manifolds. However, when the wave emerges from the acoustic field, XY planes are not iso-phase manifoldes anymore; the information about the accumulated pressure along each line resides in the phase of the emerging radiation, forming a phase image in the XY plane. The phase information is given by the Raman-Nath parameter:
with - the piezooptic coefficient, the optical wavelength and
the three-dimensional pressure field. The schlieren technique converts the phase information into an intensity image, detectable by a camera or a screen.

Application

The accepted gold-standard for quantitative acoustic measurement is the hydrophone. However, scanning the acoustic field with a hydrophone suffers from several limitations, giving rise to supplementary evaluation methods such as the schlieren imaging. The importance of the schlieren imaging technique is prominent in HIFU research and development.
Advantages of schlieren imaging include: