Vibrational analysis with scanning probe microscopy
The technique of vibrational analysis with scanning probe microscopy allows probing vibrational properties of materials at the submicrometer scale, and even of individual molecules. This is accomplished by integrating scanning probe microscopy and vibrational spectroscopy. This combination allows for much higher spatial resolution than can be achieved with conventional Raman/FTIR instrumentation. The technique is also nondestructive, requires non-extensive sample preparation, and provides more contrast such as intensity contrast, polarization contrast and wavelength contrast, as well as providing specific chemical information and topography images simultaneously.
History
Raman-NSOM
was described in 1984, and used in many applications since then. The combination of Raman scattering and NSOM techniques was first realized in 1995, when it was used for imaging a Rb-doped KTP crystal at a spatial resolution of 250 nm.NSOM employs two different methods for data collection and analysis: the fiber tip aperture approach and the apertureless metal tip approach. NSOM with aperture probes has a smaller aperture that can increase the spatial resolution of NSOM; however, the transmission of light to the sample and the collection efficiency of the scattered/emitted light is also diminished. The apertureless near-field scanning microscopy was developed in the 1990s. ANSOM employs a metalized tip instead of an optical fiber probe. The performance of the ANSOM strongly depends on the electric field enhancement factor of the metalized tip. This technique is based on surface plasmon resonance which is the precursor of tip-enhanced Raman scattering and surface-enhanced Raman scattering.
In 1997, Martin and Girard demonstrated theoretically that electric field under a metallic or dielectric tip can be strongly enhanced if the incident field is along the tip axis. Since then a few groups have reported Raman or fluorescence enhancement in near field optical spectroscopy by apertureless microscopy. In 2000, T. Kalkbrenner et al. used a single gold particle as a probe for apertureless scanning and presented images of an aluminium film with 3 μm holes on a glass substrate. The resolution of this apertureless method was 100 nm, that is comparable to that of fiber-based systems Recently, a carbon nanotube having a conical end, tagged with gold nanoparticles, was applied as a nanometer-resolution optical probe tip for NSOM. NSOM images were obtained with a spatial resolution of ~5 nm, demonstrating the potential of a composite CNT probe tip for nanoscale-resolution optical imaging.
Tip-enhanced Raman scattering
There are two options for realizing apertureless NSOM-Raman technique: TERS and SERS. TERS is frequently used for apertureless NSOM-Raman and can significantly enhance the spatial resolution. This technique requires a metal tip to enhance the signal of the sample. That is why an AFM metal tip is usually used for enhancing the electric field for molecule excitation. Raman spectroscopy was combined with AFM in 1999. A very narrow aperture of the tip was required to obtain a relatively high spatial resolution; such aperture reduced the signal and was difficult to prepare. In 2000, Stȍckle et al. first designed a setup combining apertureless NSOM, Raman and AFM techniques, in which the tip had a 20 nm thick granular silver film on it. They reported a large gain in the Raman scattering intensity of a dye film deposited on a glass substrate if a metal-coated AFM tip was brought very close to the sample. About 2000-fold enhancement of Raman scattering and a spatial resolution of ~55 nm were achieved.Similarly, Nieman et al. used an illuminated AFM tip coated with a 100 nm thick film of gold to enhance Raman scattering from polymers samples and achieved a resolution of 100 nm. In the early research of TERS, the most commonly used coating materials for the tip probe were silver and gold. High-resolution spatial maps of Raman signals were obtained with this technique from molecular films of such compounds as brilliant cresyl blue, malachite green isothiocyanate and rhodamine 6G, as well as individual carbon nanotubes.
IR-NSOM and AFM
IR near-field scanning optical microscopy is a powerful spectroscopic tool because it allows subwavelength resolution in IR spectroscopy. Previously, IR-NSOM was realized by applying a solid immersion lens with a refractive index of n, which shortens wavelength to, compared to FTIR-based IR microscopy. In 2004, an IR-SNOM achieved a spatial resolution ~λ/7 that is less than 1 μm. This resolution was further improved to about λ/60 that is 50–150 nm for a boron nitride thin film sample.IR-NSOM uses an AFM to detect the absorption response of a material to the modulated infrared radiation from an FTIR spectrometer and therefore is also referred to as AFM/FTIR spectroscopy. Two approaches have been used to measure the response of polymer systems to infrared absorption. The first mode relies on the AFM contact mode, and the second mode of operation employs a scanning thermal microscopy probe to measure the polymer's temperature increase. In 2007, AFM was combined with infrared attenuated total reflection spectroscopy to study the dissolution process of urea in a cyclohexane/butanol solution with a high spatial resolution.
Theory and instrumentation
Raman-NSOM
There are two modes for the operation of NSOM technique, with and without an aperture. These two mode have also been combined with the near-field Raman spectroscopy. The near-field aperture must be nanosized that complicates the probe manufacturing process. Also, the aperture method usually has a very weak signal due to weak excitation and Raman scattering signal. Overall, these factors lower the signal-to-noise ratio in aperature based NSOM/Raman technique. Apertureless probes are based on a metal-coated tip and provide a stronger signal.Aperture-based detection
Although the apertureless mode is more promising than the aperture mode, the latter is more widely used because of easier instrumental setup and operation. To obtain a high resolution Raman micrograph/spectrum, the following conditions should be met: the size of the aperture must be on the order of the wavelength of the excitation light. The distance from the tip of the probe to the sample must be smaller than excitation wavelength. The instrument must remain stable over a long time. An important AFM feature is the ability to accurately control the distance between the sample and probe tip, which is the reason why the AFM-Raman combination is preferred for realizing Raman-NSOM.Apertureless mode
The main drawback of the aperture mode is that the small aperture size reduces the signal intensity and is difficult to fabricate. Recently, researchers have focused on the apertureless mode, which utilizes SPR theory to produce stronger signals. There are two techniques supporting this mode: SERS and TERS.TERS technique
Theory and instrumentation of Raman/AFM and IR/AFM combine the theory of SPR and Raman scattering, and this combination is based on TERS. In TERS, the electric field of excitation source induces an SPR in the tip of the probe. If the electric field vector of the incidence light is perpendicular to the metal tip axis, the free electrons are driven to the sides lateral of the tip. If it is parallel to the tip axis, the free electrons on the surface of the metal are confined to the end of the apex of tip. As a consequence, there is a very large electric-field enhancement which is sensed by the molecules close to it leading to a stronger signal.A typical approach in a TERS experiment is to focus the laser beam on a metal tip with the light polarized along the tip axis, followed by collection of the surface-enhanced Raman scattered light from the sample in the enhancement zone of the tip using optics.
Depending on the sample and experiment, different illumination geometries have been applied in TERS experiments, as shown in figure 4. With p-polarized incidence light, the plasmon excitation at the tip is most efficient. If the focusing objective lens is also used for collecting the scattered photons, the optimum angle is around 55° with respect to the surface normal. This is because the scattering lobe is maximum with this configuration and it provides a much enhanced signal. The setup of figure 4 is usually used for the large thick samples. Setup handles semi-transparent or transparent samples, such as single cells, tissue samples and biopolymers. The setup of figure 4 is preferred for opaque samples because all the light would be focused by the parabolic mirror.
Figure 2. S- and p-polarization with laser excitation. | Figure 3. TERS setup using the back-excitation mode. | Figure 4. TERS setup: side illumination ; in-line illumination ; in-line illumination. |
Comparison of TERS and SERS
Both TERS and SERS rely on a localized surface plasmon for increasing the ought-to-be weak Raman signal. The only difference between them is that the sample in SERS has a rough surface that hinders application of a sharp AFM-like tip. TERS, on the other hand, uses a metal-coated tip having some roughness at nanoscale. The “hot spot” theory is very popular in explaining the large enhancement in the signal. That is, the signal from “hot spots” on the surface of the sample dominates the total signal from the sample. This is also reinforced by the fact that the distance between nanoparticles and sample is an important factor in obtaining high Raman signal.Raman/AFM instrumentation
The Raman/AFM technique has two approaches: aperture and apertureless, and the apertureless mode is realized with SERS and TERS. Figure 5 is the example of an integrated TERS system. It shows that there are five main components for a whole integrated TERS system. These components are: microscope, one objective lens, one integrated AFM head, a Raman spectrometer and a CCD. The laser is focused on the sample, on piezo-stage and the AFM tip by the moving the laser beam along the tip. The movement of the laser beam is achieved by the mirror in the top left corner. The XYZ piezo-stage in the left bottom holds the sample. In this design, the laser beam is focused on the sample through an objective lens, and the scattered light is collected by the same lens.This setup utilizes a low contact-pressure to reduce the damage to the AFM tip and sample. The laser power is typically below 1 mW. The notch filter can filter Rayleigh scattering from the excitation laser light from the back of the cantilever. The laser beam is focused on the apex of the gold-coated AFM tip and the sample. The laser scanning is completed by the moving the mirror across the approaching tip. A small enhance in background occurs when the laser spot focuses on the tip area. The movement of the XYZ piezo-stage finishes the sample scanning. The wide red signal is Raman signal which is collected through the objective lens. The same lens is also used for excitation of the sample and collecting the Raman signal.
NSOM/FTIR, AFM/FTIR and AFM-IR
Because of the diffraction limit in the resolution of conventional lens-based microscopes, namely D = 0.61λ/nsinθ, the maximum resolution obtainable with an optical microscope is ~200 nm. A new type of lens using multiple scattering of light allowed to improve the resolution to about 100 nm. Several new microscopy techniques with a sub-nanometer resolution have been developed in the last several decades, such as electron microscopy and scanning probe microscopy. SPM differs from other techniques in that the excitation and signal collection are very close to the sample. Instead of using a conventional lens to obtain magnified images of samples, an SPM scans across the sample with a very sharp probe. Whereas SEM and TEM usually require vacuum and an extensive sample preparation, SPM measurements can be performed in atmospheric or liquid conditions.Despite the achievable resolution of atomic scale for AFM and NSOM techniques, it does not provide chemical information of the sample. The infrared part of the electromagnetic spectrum covers molecular vibrations which can characterize chemical bonding within the sample.
By combining SPM and vibrational spectroscopy, AFM/IR-NSOM and AFM-IR have emerged as useful characterization tools that integrate the high spatial resolution abilities of AFM with IR spectroscopy. This new technique can be referred to as AFM/FTIR and NSOM/FTIR. AFM and NSOM can be used to detect the response when a modulated infrared radiation generated by an FTIR spectrometer is absorbed by a material. In the AFM-IR technique the absorption of the radiation by sample will cause a rapid thermal expansion wave which will be transferred to the vibrational modes of the AFM cantilever. Specifically, thermal expansion wave induces a vertical displacement of the ATM tip. A local IR absorption spectrum then can be obtained through the measurement of the amplitude of the cantilever, which is a function of the IR source wavelength. For example, when the radiation laser wavelength is tuned at the resonance frequency with the vibrational absorption frequency of the sample, the displacement intensity of the cantilever will increase until the laser wavelength reaches the maximum of sample absorption. The displacement of the cantilever will then be reduced as the laser wavelength is tuned past the absorption maximum. This approach can map chemical composition beyond the diffraction-limit resolution and can also provide three-dimensional topographic, thermal and mechanical information at the nanoscale. Overall, it overcomes the resolution limit of traditional IR spectroscopy and adds chemical and mechanical mapping to the AFM and NSOM.
Infrared light source
The ideal IR source should be monochromatic and tunable within a wide range of wavelength. According to T ∝d4/λ4, where T is the transmission coefficient, d the aperture diameter and λ is wavelength, the aperture-based NSOM/FTIR transmission is even more limited due to the long infrared wavelength; therefore, an intense IR source is needed to offset the low transmission through the optical fiber. The common bright IR light sources are the free-electron laser, color-center lasers, CO2 lasers and laser diodes. FEL is an excellent IR source, with 2–20 μm spectral range, short pulses and high average power. Alternately, a tabletop picosecond optical parametric oscillator can be used which is less expensive, but has a limited tunability and a lower power-output.NSOM/FTIR experimental setup
The essence of NSOM/FTIR is that it allows the detection of non-propagating evanescent waves in the near-field, thus yielding high spatial resolution. Depending on the detection modes of these non-propagating evanescent waves, two NSOM/FTIR instrumentations are available: apertureless NSOM/FTIR and aperture-based NSOM/FTIR.;Aperture-based NSOM/FTIR
In aperture-based NSOM/FTIR, the probe is a waveguide with a tapered tip with a very small, sub-wavelength size aperture. When the aperture is brought into the near-field, it collects the non-propagating light and guides it to the detector. In general, there are two modes when the aperture is scanned over the sample: illumination mode and collection mode.
The high-quality infrared fiber tip is very important in realizing NSOM/FTIR technique. There are several types of fibers, such as sapphire, chalcogenide glass, fluoride glass and hollow silica guides. Chalcogenide glasses are widely used because of their high transmittance in the broad IR range of 2–12 μm. The fluoride fibers also exhibit low transmitting losses beyond 3.0 μm.
;Apertureless NSOM/FTIR
The probe is a sharp metal tip ending with a single or a few atoms. The sample is illuminated from far-field and the radiation is focused at the contact area between probe and sample. When this tip approaches the sample, usually within 10 nm, the incident electromagnetic field is enhanced due to the resonant surface plasma excitation as well as due to hot-spots in the sharp tip. The dipole interaction between the tip and sample change the non-propagating waves into propagating waves by scattering, and a detector collects the signal in the far-field. An apertureless NSOM/FTIR usually has better resolution compared with aperture-based NSOM/FTIR. One main challenge in apertureless NSOM/FTIR is a strong background signal because the scattering is obtained from both near-field and remote area of the probe. Thus, the small near-field contribution to the signal has to be extracted from the background. One solution is to use a very flat sample with only optical spatial fluctuation. Another solution is to apply constant-height mode scanning or pseudo-constant-height mode scanning.
;Experimental scheme of aperture-based NSOM/FTIR
Figure 8 shows the experimental setup used in NSOM/FTIR in the external reflection mode. FEL source is focused on the sample from the far-field using a mirror. The distance between the probe and a sample is kept at a few nanometers during scanning.
Figure 9 is the cross-section of a NSOM/FTIR instrument. As shown below, sample is placed on a piezo-electric tube scanner, in which the x-y tube has four parts, namely x+, x-, y+ and y-. Lateral oscillation of the fiber tip is induced by applying an AC voltage to a dither piezo-scanner. Also, the fiber tip is fixed to a bimorph piezo-scanner so that the amplitude of the oscillation of the tip can be monitored through the scanner.
AFM/FTIR setup
;Spatial resolutionThe spatial resolution of an AFM/FTIR instrument is related to the contact area between the probe and sample. The contact area is given by a3 = 3PR/4E* and 1/E* = / E1+ / E2, where P is the force employed to the probe, n1 and n2 represent the Poisson ratios of the sample and probe, respectively, and E1 and E2 are the elastic moduli of the sample and probe materials respectively. Typically, an AFM/FTIR has a resolution of 50–400 nm, for example, 100 nm, λ/150, and λ/400.
;Instrumentation
In AFM/FIIR, an AFM probe is used to measure the absorption response of the sample to infrared radiation. The general approach for AFM/FTIR is shown in Figure 10.
There are a few different experimental setups when the infrared radiation is projected onto the sample as shown below: top, side, and bottom illumination setups.
Specifically, a sample is mounted onto an infrared-transparent zinc selenide prism for excitation purposes, then an optical parametric oscillator -based tunable IR lased is radiated on the molecules to be probed by the instrument. Similar to conventional ATR spectroscopy, IR beam illuminates the sample through total internal reflection mechanism. The sample will heat up while absorbing radiation which causes a rapid thermal expansion of the sample surface. This expansion will increase the resonant oscillations of the AFM cantilever in a characteristic ringdown pattern. Through Fourier transformation analysis, the signal could be isolated to obtain the amplitudes and frequencies of the oscillations. The amplitudes of the cantilever provide information of local absorption spectra, whereas the oscillation frequencies depend on the mechanical stiffness of the sample.
Pros and cons
NSOM combined with FTIR/Raman technisues can provide local chemical information together with topographical details. This technique is non-destructive and can work in a variety of environments, for example, when detecting single biomolecules. The illuminated area of sample is relatively big at around 1 μm. However, the sampling area is only ~10 nm. This means that a strong background from an unclean tip contributes to the overall signal, hindering the signal analysis.The Raman spectroscopy in general could be time-consuming due to the low scattering efficiency. It usually takes several minutes to accumulate a conventional Raman spectrum, and this time could be much longer in Raman-NSOM; for example, 9 hours for a 32×32-pixel image. As to near-field IR/AFM, high optical losses in aqueous environments reduces the signal-to-noise ratio.
Applications
Improving the resolution and enhancing the instrumentation with user-friendly hardware and software will make AFM/NSOM coupled with IR/Raman a useful characterization tool in many areas including biomedical, materials and life sciences. For example, this technique was used in detecting the spin-cast thin film of poly with polystyrene on it by scanning the tip over the sample. The shape and size of polystyrene fragments was detected at a high spatial resolution due to its high absorption at specific resonance frequencies. Other examples include inorganic boron nitride thin films characterization with IR-NSOM. The images of single molecule rhodamine 6G was obtained with a spatial resolution of 50 nm. These techniques can be also used in numerous biological related applications including the analysis of plant materials, bone, and single cells. Biological application was demonstrated by detecting details of conformation changes of cholesteryl-oleate caused by FEL irradiation with a spatial resolution below the diffraction limit. Researchers also used Raman/NSOM in tracking the formation of energy-storing polymer polyhydroxybutyrate in bacteria Rhodobacter capsulatus.This characterization tool may also help in the kinetic studies on physical and chemical processes at a wide variety of surfaces giving chemical specificity via IR spectroscopy as well as high-resolution imaging via AFM. For example, the study of the hydrogen termination of Si surface was performed by studying the absorbance of Si-O bond to characterize the reaction between silicon surface and atmospheric oxygen. Studies were also conducted where the reactivity of a polymer, a 1000-nm-thick poly- combined with a photochemically modified 500-nm-thick poly, with water vapor depicted the different absorption bands before and after water uptake by the polymer. Not only the increased swell of PMAA was observed but also the different absorption ability of water was shown by the different transmission of IR light at a much smaller dimension. These results are related to polymer, chemical and biological sensors, and tissue engineering and artificial organ studies. Because of their high spatial resolution, NSOM/AFM-Raman/IR techniques can be used for measuring the width of multilayer films, including layers which are too small to be probed with conventional IR or Raman spectroscopy.